High pressure gas cells

High Pressure Cells

High Pressure Cells
View full-size image

The Pressure and Furnace section have a diverse range of cells and pressure systems, for both gas and liquid applications.

The cells shown below are examples of the high pressure cells available at ISIS. The titanium Zirconium (TiZr) cells are used on the diffraction type instruments and the 7075 aluminum alloy cells are used for spectroscopy experiments. These cells can be used for powder, liquid and crystal samples.

The photograph shows non standard variations of the other TiZr and aluminum 7075 alloy cells shown. The working temperature range of the cells is from 2K to 300K.

Material

Max working pressure at room temperature

Sample volume (approx)
and sample space dimensions

Other info

TiZr

5,600bar

1.54cc
Ø7mm x 40mm

No Hydrogen
Null scattering alloy

TiZr5,000barFlat geometry with 6 vertical holesN/A
BeCu5,000barØ3mm x 12mmFor use in SQUID

7075 Al Alloy

4,500bar

1.54cc
Ø7mm x 40mm

N/A

Inconel7,360bar1.54cc
Ø7mm x 40mm
N/A
Be/Cu7,360bar1.54cc
Ø7mm x 40mm
Hydrogen compatible
Be/Cu and TiZr two layer cell8000 bar1.54cc
Ø7mm x 40mm
Hydrogen compatible outer layer null scattering alloy
The cells named above only represent a small portion of what we have at ISIS. If you have any specific requirements or questions please contact Chris Goodway.

The cells shown above are examples of the high pressure cells available at ISIS. The titanium Zirconium (TiZr) cells are used on the diffraction type instruments and the 7075 aluminum alloy cells are used for spectroscopy experiments. These cells can be used for powder, liquid and crystal samples.

Bookmark and Share
Skip to the top of the page