XRD technical information

Summary of XRD setup

The XRD is typically set up with a silver anode tube and a scintillation detector.  Experiments with a sample of SiO2 have demonstrated that the statistical error across the angular range varies greatly, and good data typically requires 24 hours to collect.  

Scan times of this length are made on the assumption that the signal is sufficiently good; a range of capillary sizes, from 0.5 to 2.5 mm, are available to allow the balance between scattering intensity and absorption to the optimised.

For experiments which do not need high Q data (Qmax< 17 Å-1) measurements can be made with a Mo tube and an X'Celerator detector for faster counting times.   Data collection using this technique is still under development.

The low temperature stage has not yet been commissioned.  If you are interested in using this, please contact us to discuss your experiment.

 

Summary of XRD settings

SettingsDescription
TubeAg or Mo anodes available
DetectorsScintillation detector for Ag tube
X'Celerator or scintillation detector for Mo tube
PowerTypically run at full power (60kV/35mA)
StageCapillary stage with option for flat plate
Angular/Q range3.6 - 156o/ possible Qmax in excess of 20 Å-1.
EnvironmentPotential to measure samples over a range of temperatures, down to -70oC, using a cryostream